ESS-2000AX |
Electrostatic Discharge Simulator
ESS-2000AXCONFORMING TO IEC 61000-4-2ESD Generator's brochure ESS. PDF(2362KB) for additional product information.
Get free"FAQ for Electro-Static Discharging Test"by answering a brief questionnaire. For more information.....
<Limited Time Offer>
Buy NoiseKens ESS-2000AX and get free"ESD Testing Manual."
Manual SampleEnglish Chinese
TC-815R
TC-815R ESS-2000AX
(The Gun stand PS-806 in the photo is an optional essory.)
GENERAL
The ESS-2000AX ESD Simulator generates highly consistent, repeatable ESD pulses up to 30kV in both contact and air discharge modes. Compliant with IEC61000-4-2 standard, the unit features a 5.7-inch display, a lightweight gun and optic serial interface. A variety of capacitor and resistor units are available and are easily changeable within the gun. The fully- programmable, menu-driven simulator offers three trigger modes.
FEATURES
Fully programmable menu-driven simulator providing three operation modes: IEC standard, Manual and SequenceA 5.7-inch color LCD for easy setting and good visibilityUnique shape for the operator's easy ess to the control and displays even when the unit is put on the floor level (ground plane)Communications with PC through the optic serial interfaceA wide variety of the dedicated options
SPECIFICATIONS
ItemESS-2000AX SpecificationsOutput polarityPositive or negativeOutput voltage0.20kV~30.0kV±5% (30.5kV max)
0.20kV~10.0kV: 0.01V step, 10.0kV~30.0kV:0.1V step
Discharge intervalNormal mode: 0.05s~600.0s±10%
Extra mode : 1.00s~600.0s±10%
0.05s~9.99s: 0.01s step, 10s~600.0s: 0.1s step
No. of times of discharge1~60,000 times preset on the counter or continuousDischarge modeAir/contact dischargeRadiation levelNormal/Extra modeTrigger modeGun trigger/main unit trigger/ external triggerOperation
modeIEC standardContact discharge: 2.0kV, 4.0kV, 6.0kV, 8.0kV
(4 voltages preset)
Air discharge: 2.0kV, 4.0kV, 8.0kV, 15.0kV (4 voltages preset)
ManualContact/Air discharge: 0.20kV~30.0kV
Sweep (automatic voltage ramping) available
Up to 99 units (test settings) storable
SequenceSeamless run of the selected test sequence (made of any combination of 99 units in Manual mode)
No. of units per one test sequence: up to 30 units
No. of test sequences: up to 20
Warning lampLights up while the HV circuitry is on
Blinks while discharges are in progress
Discharge gunTC-815RCharging resistance10M ohm(53 M ohm for combination with TC-815R discharge gun)AUX connectorD-SUB15 pin, female connector
Warning lamp (optional) connection
Automatic ESD Eliminator (optional) connection
External interlock input
External trigger input
Optical communication connectorOptic RS-232C for PC communicationPower supply100~240V AC 50/60 Hz ±10%Power consumption75VAOperating temperature and humidity+15~+35ºC, 25~75% (no condensation)Character displayEnglish, Chinese, Korean, JapaneseDimensions and weight260 (W) X 230 (H) X 285 (D) mm (projection excluded)
Approx. 5.3 kg
DISPLAY EXAMPLES
Display Examples in Korean Display Examples in Chinese
TOP MENUAfter pressing the main switch, press the EXIT key. This places the simulator in the initial menu, which displays three operational modes, trigger selection and utility.Manual ModeIf you desire to operate the unit in the manual mode, press the corresponding function key. The parameters to be set by the operator will appear. Discharge method (contact/air discharge), discharge voltage, number of discharges and interval can be set. Select the parameter by pressing the corresponding function key then the value can be altered by the Increment or Decrement key.Contact discharges: For contact discharge testing, after completion of required settings, press the START button and pull the trigger. The simulator will then generate the required number of pulses at the required interval. Pulling the trigger again will pause the unit. Pulling again will restart the unit.Air discharges: For air discharge testing, after completion of setting, press the START key. To carry out air discharges, pull and hold the trigger to maintain the HV relay in the on status and approach the discharge tip to the EUT.When the SWEEP function is active, the simulator generates discharges in an automatic voltage ramp. The initial, final and step voltages can be freely set. In this mode, the number of discharges set is that in each stepFor example, when the simulator is set to 5kV for the initial voltage, 10kV for final voltage and 1kV for step voltage, in a way of 10 discharges at an interval of 1 second, it produces 10 pulses at 5kV at an interval of 1 second and proceeds to 6kV pulses, also 10 discharges. These steps continue until the simulator has completed 10 pulses of 10kV.Two different ways of pulling the trigger: When the trigger is pulled and then released quickly, the simulator operates in a way that it pauses before it proceeds to next step voltage. For continuous operation, pull the trigger for more than 2 seconds. A sign of "CONTINUE" is indicated on the screen and the simulator runs a complete step ramp test without further operator’s trigger.IEC Standard ModeThe four test levels are preset. Select the desired level and run a test.Sequence ModeThe simulator can store up to 99 tests including Sweep settings. Each test is given a number (called Unit No). Any combination of units selected from 99 tests in the Manual Mode can consist of one test sequence having 30 units maximum.Twenty test sequences can be stored. For a unit setting, press EDIT button. Settings of voltage, etc. can be done in the same way as in the Manual mode.
OPTIONS
Automatic ESD Eliminator 01-00013AConnected to the ESS-2000AX simulator main unit, the Automatic ESD Eliminator automatically removes residual charges on the EUT. A control signal from the ESS-2000AX turns on the switch of the eliminator to drain away the residual charges to the ground. Two operation modes are provided: turn-on after each discharge and turn-on after preset number of discharges (Not compliant with the standard).Gun holder 03-00056AA gun holder that can be screwed to the left side of ESS-2000AX. 03-00056A Usage example Warning lamp 11-00014A AUX connector junction box 05-00052AWarning Lamp and Automatic ESD Eliminator options and external trigger circuit can be simultaneously connected.Optical USB module 07-00022AOptical conversion adapter to be free from the EM interference when communicating with PC.Optical RS232 module 07-00017AOptical conversion adapter to be free from the EM interference when communicating with PC